陆艳艳,何世堂,孙林,朱宏伟,任 伟,刘久玲.快速声表面波气相色谱技术检测水中苯系物*[J].,2020,39(3):325-328 |
快速声表面波气相色谱技术检测水中苯系物* |
Determination of benzene series in water by using the fast SAW gas chromatograph |
投稿时间:2019-07-15 修订日期:2020-04-30 |
中文摘要: |
建立了便携式声表面波气相色谱(GC-SAW)快速检测苯系物方法。在该条件下顶空测试: 检测器温度25℃,DB-5毛细管色谱柱:初始温度40℃,10℃/s程序升温至180℃。结果表明,氯苯、间二甲苯、苯乙烯、异丙苯保留时间分别为1.72、1.86、2.00、2.22s,检出限分别为:1.34、0.55、0.50、0.39 μg/L,测得RSD值为4.08%-5.98%,回收率为 82.7%-103.8%,且线性良好,线性相关系数为0.9946-0.9993。因此,该方法适用于水中苯系物的快速应急监测。 |
英文摘要: |
A method of benzene series rapid detection has been developed using gas chromatography with surface acoustic wave detector (GC-SAW). Benzene series was tested by headspaceSunder this condition. The SAW detector temperature was set at 25℃. The initial temperature of DB-5 capillary column was 40℃, and was temperature-programmed to 180℃ with the rate of 10℃/s. The results show that the retention times of chlorobenzene, m-xylene, styrene and cumene were 1.72, 1.86, 2.00, 2.22s respectively, the detection limits were 1.34, 0.55, 0.50, 0.39 μg/L respectively, and the RSD were in 4.08%-5.98%. The recoveries were 82.7%-103.8%, and the linearities were good,linear correlation coefficients were 0.9946-0.9993. Therefore, this method is suitable for benzene series rapidly emergency monitoring in water. |
DOI:10.11684/j.issn.1000-310X.2020.03.001 |
中文关键词: 声表面波气相色谱 顶空进样 苯系物 快速检测 |
英文关键词: GC-SAW headspace injection benzene series rapidly detect |
基金项目: |
|
摘要点击次数: 1767 |
全文下载次数: 1582 |
查看全文
查看/发表评论 下载PDF阅读器 |